Abstract
Major contributing parameters to hip implant dislocation include preoperative, intra-operative and post-operative factors. Implant geometry are design as well as non-design related. Femoral and acetabular component design features causing dislocation and/or resisting it are elucidated. Twelve implants were designed during this investigation were analyzed for dislocation resistance. A safe zone, establishes combinations of implant dimensions, was analyzed for all the 12 implants where implants were dislocation resistant. Head diameters between 26 mm to 32 mm, neck diameters closer to 14 mm, and neck angle between 25 to 35º were examined to be the safest ranges for hip implant designs.